This subsystem defines all channel settings for a Segmented Frequency sweep. A segmented sweep is activated via [SENSe<Ch>:]SWEep:TPYE SEGMent.
Command tree: [SENSe<Ch>:]SEGMent<Seg>
:BWIDth
:FREQuency
:POWer
:SWEep
The commands in the [SENSe<Ch>:]SEGMent<Seg>...
subsystem do not accept the step parameters UP and DOWN. Numeric values
can be entered directly or using the DEFault, MINimum, MAXimum parameters.
Inserts a new sweep segment using default channel settings (Insert New Segment). The added segment covers the frequency interval between the maximum frequency of the existing sweep segments and the stop frequency of the entire sweep range.
Use [SENSe<Ch>:]SEGMent<Seg>:INSert
to create a segment with specific channel settings.
|
<Ch> |
Channel number. If unspecified the numeric suffix is set to 1.
|
|
<Seg> |
Sweep segment number. Segment numbers must be sequential. If n segments already exist, the added segment must have the segment number n+1.
|
|
*RST value |
–
|
|
SCPI, Command Types |
Device-specific, no query.
|
|
Example: |
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ Create a sweep segment with a sweep range between 1.0 MHz and 1.5 MHz. SEGM2:ADD Create a second sweep segment. The frequency range of the second segment will be between 1.5 MHz and the maximum frequency of the analyzer.
|
Defines the resolution bandwidth of the analyzer (Meas. Bandwidth) in sweep segment no. <Seg>. At the same time, the command activates separate bandwidth setting in all sweep segments ([SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:CONTrol ON).
|
<Ch> |
Channel number 1.
|
|
<Seg> |
Sweep segment number.
|
|
<numeric_value> |
Internal source power. |
|
Range [def. unit] |
1.0E-6 Hz to 5 MHz [Hz]. UP and DOWN increment/decrement the bandwidth in 1-2-5 steps for each decade. The analyzer rounds up any entered value between these steps and rounds down values exceeding the maximum bandwidth. |
|
*RST value |
10 kHz
|
|
SCPI, Command Types |
Device-specific, command or query
|
|
Example: |
SEGM:ADD Create a new sweep segment no. 1 in channel no. 1 using default settings and thus –10 dBm internal source power. SEGM:BWID 1 MHZ Increase the resolution bandwidth to 1 MHz.
|
Qualifies whether or not the Meas. Bandwidth can be set independently for each sweep segment.
|
<Ch> |
Channel number 1.
|
|
<Seg> |
Sweep segment number.
|
|
<Boolean> |
ON: The bandwidth can be set independently for each sweep
segment. |
|
*RST value |
OFF. The parameter is automatically switched to ON when a bandwidth is entered using [SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution].
|
|
SCPI, Command Types |
Device-specific, command or query
|
|
Example: |
SEGM:ADD Create a new sweep segment no. 1 in channel no. 1 using default settings and thus 10 kHz measurement bandwidth. SEGM:BWID 1 MHZ Increase the resolution bandwidth to 1 MHz. SEGM:BWID:CONT OFF Couple the bandwidths in all segments and reset the bandwidth in segment no. 1 to the initial value.
|
Defines the Selectivity in sweep segment no. <Seg>. At the same time, the command activates separate selectivity setting in all sweep segments ([SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:CONTrol:SELect ON).
|
<Ch> |
Channel number 1.
|
|
<Seg> |
Sweep segment number.
|
|
NORMal |
Use IF filter with normal selectivity and short settling time. |
|
HIGH |
Use IF filter with higher selectivity but larger settling time. |
|
*RST value |
NORMal
|
|
SCPI, Command Types |
Device-specific, command or query
|
|
Example: |
SEGM:ADD Create a new sweep segment no. 1 in channel no. 1 using default settings and thus –10 dBm internal source power. SEGM:BWID:SEL HIGH Activate an IF filter with high selectivity.
|
[SENSe<Ch>:]SEGMent:CLEarDeletes all sweep segments in the channel. The command is equivalent to [SENSe<Ch>:]SEGMent:DELet:ALL.
Returns the number of sweep segments in the channel including all segments that are switched off ([SENSe<Ch>:]SEGMent<Seg>[:STATe] OFF).
|
<Ch> |
Channel number.
|
|
*RST value |
–
|
|
SCPI, Command Types |
Device-specific, query only.
|
|
Example: |
SEGM:ADD Create a new sweep segment no. 1 in channel no. 1 using default settings. SEGM OFF Disable the measurement in the created sweep segment. SEGM:COUN? Query the number of segments.
|
Re-defines a sweep segment with specific channel settings (Insert New Segment). The segment replaces an existing segment <Seg> in the segment list. The modified segment must not overlap with any of the existing segments.
Entry of the first seven numeric parameters is mandatory; no default values are provided. All settings except <LO> can be changed for existing segments using other commands of the [SENSe<Ch>:]SEGMent<Seg>... subsystem.
Note: Use [SENSe<Ch>:]SEGMent<Seg>:Add to create a segment with default channel settings. Use [SENSe<Ch>:]SEGMent<Seg>:INSert (no query) to insert a new segment into the current segment list.
|
<Ch> |
Channel number.
|
|
<Seg> |
Sweep segment number. Segment numbers must be sequential. The specified segment number must be smaller or equal to the number of existing segments plus 1. If segment number <Seg> already exists, it is replaced by the new segment.
|
|
<Start>, <Stop> |
Start and stop frequency of the segment. See [SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt and [SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP. |
|
Range [def. unit] |
A new segment must not overlap with any of the existing segments. Besides the frequency range depends on the analyzer model. [Hz]
|
|
<Points> |
Number of sweep points in the segment. See [SENSe<Ch>:]SEGMent<Seg>:SWEep:POINts. |
|
Range [def. unit] |
1 to 2147483647. [1] 1 is allowed if start and stop frequencies are equal.
|
|
<Power> |
Internal source power in the segment. See [SENSe<Ch>:]SEGMent<Seg>:POWer. |
|
Range [def. unit] |
–40 dBm to +10 dBm. The exact range depends on the analyzer model; refer to the data sheet [dBm]. UP and DOWN increment/decrement the source power in 1-dB steps.
|
|
<Time> |
Duration of the sweep in the segment. See [SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME. In the setting [SENSe<Ch>:]SEGMent<Seg>:INSert:SELect DWELl, this parameter is replaced by <Point Delay>. |
|
Range [def. unit] |
Depending on other channel settings [s]. AUTO activates automatic sweep time setting in the segment, which is equivalent to the minimum sweep time possible.
|
|
<Point Delay> |
Delay for each partial measurement in the segment. See [SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl. In the setting [SENSe<Ch>:]SEGMent<Seg>:INSert:SELect SWTime, this parameter is replaced by <Time>. |
|
Range [def. unit] |
0 s to 2.5E+003 s. [s] AUTO activates automatic sweep time setting in the segment, which is equivalent to a meas. delay of 0 s.
|
|
|
Ignored parameter, should be set to the default value 0. |
|
<Meas. Bandwidth> |
Resolution bandwidth in the segment. See [SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]. |
|
Range [def. unit] |
1.0E-6 Hz to 5 MHz [Hz].
|
|
<LO> |
Position of the local oscillator frequency LO relative to the RF frequency (Spur Avoid). In remote control this parameter must be set when a sweep segment is created. |
|
Range [def. unit] |
POSitive: LO > RF
|
|
<Selectivity> |
Selectivity of the IF filter. See [SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect. |
|
Range [def. unit] |
NORMal | HIGH [–]
|
|
SCPI, Command Types |
Device-specific, with query (used to retrieve the channel settings for a particular sweep segment).
|
|
Example: |
SEGM:ADD Create a new sweep segment no. 1 in channel no. 1 using default settings. SEGM:DEF? Query the channel settings for the new segment.
|
Defines whether the sweep time of a new segment, i.e. numeric parameter no. 5 of the command [SENSe<Ch>:]SEGMent<Seg>:DEFine, is entered as a segment sweep time or as a meas. delay.
|
<Ch> |
Channel number 1.
|
|
<Seg> |
Sweep segment number.
|
|
SWTime |
Use segment sweep time. |
|
DWELl |
Use meas. delay. |
|
*RST value |
SWTime
|
|
SCPI, Command Types |
Device-specific, command or query.
|
|
Example: |
SEGM:DEF:SEL DWEL Select the meas. delay to determine the sweep time in a new sweep segment. SEGM:DEF 1MHZ, 1.5MHZ, 111, -21DBM, 0.01S, 0, 10KHZ Create a sweep segment with a sweep range between 1.0 MHz and 1.5 MHz and a meas. delay of 10 ms.. SEGM:SWE:TIME? Query the sweep time in the new segment.
|
Deletes the specified (single) sweep segment (Del. Selected Segment). [SENSe<Ch>:]SEGMent<Seg>:DELete:ALL deletes all segments in the channel.
|
<Ch> |
Channel number.
|
|
<Seg> |
Sweep segment number. If unspecified, the value is set to 1.
|
|
*RST value |
–
|
|
SCPI, Command Types |
Device-specific, no query.
|
|
Example: |
SEGM:ADD Create a new sweep segment no. 1 in channel no. 1 using default settings and thus NORMal selectivity. SEGM:DEL Delete the created segment.
|
Deletes all sweep segments in the channel (Del. All Segments). [SENSe<Ch>:]SEGMent<Seg>:DELete deletes a single segment.
|
<Ch> |
Channel number.
|
|
*RST value |
–
|
|
SCPI, Command Types |
Device-specific, no query.
|
|
Example: |
SEGM:ADD Create a new sweep segment no. 1 in channel no. 1 using default settings and thus NORMal selectivity. SEGM:ALL Delete the created segment and all segments in the channel created before.
|
Returns the center frequency of sweep segment no. <Seg>.
|
<Ch> |
Channel number.
|
|
<Seg> |
Sweep segment number.
|
|
Response |
Center frequency of the sweep. |
|
Range [def. unit] |
Depending on the instrument model [Hz]. |
|
*RST value |
–
|
|
SCPI, Command Types |
Device specific, query only.
|
|
Example: |
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ Create a sweep segment with a sweep range between 1.0 MHz and 1.5 MHz. SEGM:FREQ:CENT? Query the center frequency of the created segment. The response is 1.2500000000E+006.
|
Note: The frequency range of the sweep segment can be changed via [SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt and [SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP.
Returns the width of the frequency range of sweep segment no. <Seg>.
|
<Ch> |
Channel number.
|
|
<Seg> |
Sweep segment number.
|
|
Response |
Frequency span of the sweep. |
|
Range [def. unit] |
Depending on the instrument model [Hz]. |
|
*RST value |
–
|
|
SCPI, Command Types |
Device specific, query only.
|
|
Example: |
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ Create a sweep segment with a sweep range between 1.0 MHz and 1.5 MHz. SEGM:FREQ:SPAN? Query the span of the created segment. The response is 5.0000000000E+005.
|
Note: The frequency range of the sweep segment can be changed via [SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt and [SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP.
Defines the Start frequency of sweep segment no. <Seg>.
|
<Ch> |
Channel number.
|
|
<Seg> |
Sweep segment number.
|
|
<numeric_value> |
Start frequency of the sweep. |
|
Range [def. unit] |
Depending on the instrument model [Hz]. The increment is 0.1 kHz. |
|
*RST value |
– (no sweep segment defined after reset)
|
|
SCPI, Command Types |
Device-specific, command or query
|
|
Example: |
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ Create a sweep segment with a sweep range between 1.0 MHz and 1.5 MHz. SEGM:FREQ:STAR? Query the start frequency of the created segment. The response is 1.0000000000E+006.
|
Note: If the start frequency entered is greater than the current stop frequency ([SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP), the stop frequency is set to the start frequency plus the minimum frequency span of 1 Hz.
Defines the Stop frequency of sweep segment no. <Seg>.
|
<Ch> |
Channel number.
|
|
<Seg> |
Sweep segment number.
|
|
<numeric_value> |
Stop frequency of the sweep. |
|
Range [def. unit] |
Depending on the instrument model [Hz]. The increment is 0.1 kHz. |
|
*RST value |
– (no sweep segment defined after reset)
|
|
SCPI, Command Types |
Device-specific, command or query
|
|
Example: |
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ Create a sweep segment with a sweep range between 1.0 MHz and 1.5 MHz. SEGM:FREQ:STOP? Query the stop frequency of the created segment. The response is 1.5000000000E+006.
|
Note: If the stop frequency entered is smaller than the current start frequency ([SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt), the start frequency is set to the stop frequency minus the minimum frequency span ([SENSe<Ch>:]SEGMent<Seg>:FREQuency:SPAN).
Inserts a new sweep segment with specific channel settings (Insert New Segment). The new segment must not overlap with any of the existing segments.
Entry of the first seven numeric parameters is mandatory; no default values are provided. All settings except <LO> can be changed for existing segments using other commands of the [SENSe<Ch>:]SEGMent<Seg>... subsystem.
Note: Use [SENSe<Ch>:]SEGMent<Seg>:Add to create a segment with default channel settings. Use [SENSe<Ch>:]SEGMent<Seg>:DEFine to change or query all settings of an existing segment.
|
<Ch> |
Channel number.
|
|
<Seg> |
Sweep segment number. Segment numbers must be sequential. The specified segment number must be smaller or equal to the number of existing segments plus 1. Moreover, segment numbers must selected such that the corresponding frequency ranges are in ascending order. If one or more sweep segments with segment numbers <Seg> or larger exist in the current channel, then all these existing segment numbers are incremented by 1 and the new segment is inserted as segment no. <Seg>.
|
|
<Start>, <Stop> |
Start and stop frequency of the segment. See [SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt and [SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP. |
|
Range [def. unit] |
A new segment must not overlap with any of the existing segments. Besides the frequency range depends on the analyzer model. [Hz]
|
|
<Points> |
Number of sweep points in the segment. See [SENSe<Ch>:]SEGMent<Seg>:SWEep:POINts. |
|
Range [def. unit] |
1 to 2147483647. [1] 1 is allowed if start and stop frequencies are equal.
|
|
<Power> |
Internal source power in the segment. See [SENSe<Ch>:]SEGMent<Seg>:POWer. |
|
Range [def. unit] |
–40 dBm to +10 dBm. The exact range depends on the analyzer model; refer to the data sheet [dBm]. UP and DOWN increment/decrement the source power in 1-dB steps.
|
|
<Time> |
Duration of the sweep in the segment. See [SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME. In the setting [SENSe<Ch>:]SEGMent<Seg>:INSert:SELect DWELl, this parameter is replaced by <Point Delay>. AUTO activates automatic sweep time setting in the segment, which is equivalent to the minimum sweep time possible. |
|
Range [def. unit] |
Depending on other channel settings [s].
|
|
<Point Delay> |
Delay for each partial measurement in the segment. See [SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl. In the setting [SENSe<Ch>:]SEGMent<Seg>:INSert:SELec SWTime, this parameter is replaced by <Time>. |
|
Range [def. unit] |
0 s to 2.5E+003 s. [s] AUTO activates automatic sweep time setting in the segment, which is equivalent to a meas. delay of 0 s.
|
|
|
Ignored parameter, should be set to the default value 0. |
|
<Meas. Bandwidth> |
Resolution bandwidth in the segment. See [SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]. |
|
Range [def. unit] |
1.0E-6 Hz to 5 MHz [Hz].
|
|
<LO> |
Position of the local oscillator frequency LO relative to the RF frequency (Spur Avoid). In remote control this parameter must be set when a sweep segment is created. |
|
Range [def. unit] |
POSitive: LO > RF
|
|
<Selectivity> |
Selectivity of the IF filter. See [SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect. |
|
Range [def. unit] |
NORMal | HIGH [–]
|
|
<IF Gain> |
IF gain in the receiver path, valid for all ports. |
|
Range [def. unit] |
LNOise: Low Noise, for high dynamic range (low input levels)
|
|
SCPI, Command Types |
Device-specific, no query.
|
|
Example: |
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ Create a sweep segment with a sweep range between 1.0 MHz and 1.5 MHz. SEGM2:ADD Create a second sweep segment. The frequency range of the second segment will be between 1.5 MHz and the maximum frequency of the analyzer.
|
Defines whether the sweep time of a new segment, i.e. numeric parameter no. 5 of the command [SENSe<Ch>:]SEGMent<Seg>:INSert, is entered as a segment sweep time or as a meas. delay.
|
<Ch> |
Channel number 1.
|
|
<Seg> |
Sweep segment number.
|
|
SWTime |
Use segment sweep time. |
|
DWELl |
Use meas delay. |
|
*RST value |
SWTime
|
|
SCPI, Command Types |
Device-specific, command or query
|
|
Example: |
SEGM:INS:SEL DWEL Select the meas. delay to determine the sweep time in a new sweep segment. SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.01S, 0, 10KHZ Create a sweep segment with a sweep range between 1.0 MHz and 1.5 MHz and a meas. delay of 10 ms.. SEGM:SWE:TIME? Query the sweep time in the new segment.
|
[SENSe<Ch>:]SEGMent<Seg>:OVERlap
<Boolean>Queries whether the analyzer supports overlapping sweep segments.
|
<Ch> |
Channel number 1.
|
|
<Seg> |
Sweep segment number.
|
|
<Boolean> |
ON | OFF: No effect. |
|
*RST value |
OFF. If used as a query, the command returns the information that overlapping sweep segments are not supported (OFF).
|
|
SCPI, Command Types |
Device-specific, command (no effect) or query
|
Defines the Power of the internal signal source in sweep segment no. <Seg>. At the same time, the command activates separate power control in all sweep segments ([SENSe<Ch>:]SEGMent<Seg>:POWer[:Level]:CONTrol ON).
|
<Ch> |
Channel number 1.
|
|
<Seg> |
Sweep segment number.
|
|
<numeric_value> |
Internal source power. |
|
Range [def. unit] |
–40 dBm to +10 dBm. The exact range depends on the analyzer model; refer to the data sheet [dBm]. |
|
*RST value |
–10 dBm.
|
|
SCPI, Command Types |
Device-specific, command or query
|
|
Example: |
SEGM:ADD Create a new sweep segment no. 1 in channel no. 1 using default settings and thus –10 dBm internal source power. SEGM:POW -20 Decrease the power to –20 dBm.
|
Qualifies whether or not the Power of the internal signal source can be set independently for each sweep segment.
|
<Ch> |
Channel number 1.
|
|
<Seg> |
Sweep segment number.
|
|
<Boolean> |
ON: The power can be set independently for each sweep
segment. |
|
*RST value |
OFF. The parameter is automatically switched to ON when a segment level is entered using [SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel].
|
|
SCPI, Command Types |
Device-specific, command or query
|
|
Example: |
SEGM:ADD Create a new sweep segment no. 1 in channel no. 1 using default settings and thus –10 dBm internal source power. SEGM:POW -20 Decrease the power to –20 dBm. SEGM:POW:CONT OFF Couple the powers in all segments and reset the power in segment no. 1 to the initial value.
|
Activates or deactivates the sweep segment <Seg>. Sweep points belonging to inactive segments only are not measured
|
<Ch> |
Channel number. If unspecified the numeric suffix is set to 1.
|
|
<Seg> |
Sweep segment number.
|
|
<Boolean> |
ON | OFF - Activates or deactivates the measurement in sweep segment <Seg>. |
|
*RST value |
ON
|
|
SCPI, Command Types |
Device-specific, command or query
|
|
Example: |
SEGM:ADD Create a new sweep segment no. 1 in channel no. 1 using default settings. SEGM OFF Disable the measurement in the created sweep segment.
|
Defines the total number of measurement Points in sweep segment no. <Seg>.
|
<Ch> |
Channel number 1.
|
|
<Seg> |
Sweep segment number.
|
|
<numeric_value> |
Number of points in the segment. |
|
Range [def. unit] |
1 to 2147483647. [1] 1 is allowed if start and stop frequencies are equal. |
|
*RST value |
51
|
|
SCPI, Command Types |
Device-specific, command or query
|
|
Example: |
SEGM:ADD Create a new sweep segment no. 1 in channel no. 1 using default settings and thus 51 sweep points. SEGM:SWE:POIN 401 Increase the number of points to 401.
|
Sets the duration of the sweep in sweep segment no. <Seg> (Segment Sweep Time). At the same time, the command activates separate sweep time setting in all sweep segments ([SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME:CONTrol ON).
|
<Ch> |
Channel number 1.
|
|
<Seg> |
Sweep segment number.
|
|
<numeric_value> |
Internal source power. |
|
Range [def. unit] |
The minimum duration depends on the other channel settings, in particular on the number of points ([SENSe<Ch>:]SEGMent<Seg>:SWEep:POINts), the IF bandwidth (BWID) and the delay for each partial measurement ([SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl). The maximum is 1000 s. [s] Changing the duration leaves the number of points unchanged but directly affects the delay. |
|
*RST value |
MIN, depending on the channel settings. This default value corresponds to automatic sweep time setting in manual control.
|
|
SCPI, Command Types |
Device-specific, command or query
|
|
Example: |
SEGM:ADD Create a new sweep segment no. 1 in channel no. 1 using default settings and thus –10 dBm internal source power. SEGM:SWE:TIME 0.1 Increase the segment sweep time to 0.1 s.
|
Qualifies whether or not the Segment Sweep Time can be set independently for each sweep segment.
|
<Ch> |
Channel number 1.
|
|
<Seg> |
Sweep segment number.
|
|
<Boolean> |
ON: The time can be set independently for each sweep segment.
|
|
*RST value |
OFF. The parameter is automatically switched to ON when a segment sweep time is entered using [SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME or if the channel settings in a sweep segment require a sweep time larger than the unsegmented sweep time.
|
|
SCPI, Command Types |
Device-specific, command or query
|
|
Example: |
SEGM:ADD Create a new sweep segment no. 1 in channel no. 1 using default settings and thus –10 dBm internal source power. SEGM:SWE:TIME 0.1 Increase the segment sweep time to 0.1 s. SEGM:SWE:TIME:CONT OFF Couple the sweep times in all segments and reset the sweep time in segment no. 1 to the initial value.
|
Defines the delay time for each partial measurement in sweep segment no. <Seg> (Meas. Delay). If coupling of the segments is switched on ([SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl:CONTrol ON) the delay is valid for all sweep segments in the current channel.
|
<Ch> |
Channel number.
|
|
<Seg> |
Sweep segment number.
|
|
<numeric_value> |
Meas. delay before each partial measurement. |
|
Range [def. unit] |
0 s to 2.5E+003 s. [s] Changing the delay leaves the number of points unchanged but has an impact on the duration of the sweep ([SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME). |
|
*RST value |
0 s
|
|
SCPI, Command Types |
Device-specific, command or query. This default value corresponds to automatic sweep time setting in manual control.
|
|
Example: |
SEGM:ADD Create a new sweep segment no. 1 in channel no. 1 using default settings. SEGM:SWE:DWEL 1 MS Set the meas. delay in segment no. 1 to 1 ms. SEGM:DEF? Response: 300000,8000000000,51,-300,0.056559,0,10000,POS,NORM Query the channel parameters for sweep segment 1. The response value for the segment sweep time (olive) implicitly contains the defined meas. delay.
|
Qualifies whether or not the Meas. Delay defined via [SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl can be set independently for each sweep segment.
|
<Ch> |
Channel number.
|
|
<Seg> |
Sweep segment number.
|
|
<Boolean> |
ON: The meas. delay can be set independently for each
sweep segment. |
|
*RST value |
OFF. The parameter is automatically switched to ON when a meas. delay time is entered using [SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl.
|
|
SCPI, Command Types |
Device-specific, command or query.
|
|
Example: |
SEGM:ADD Create a new sweep segment no. 1 in channel no. 1 using default settings and thus 0 s meas. delay. SEGM:SWE:DWELl 0.1 Increase the meas. delay to 0.1 s. SEGM:SWE:DWELl:CONT OFF Couple the meas. delay in all segments and reset the delay in segment no. 1 to the initial value of 0 s.
|